我想知道選擇 氧分析儀器要知道什麽? 要詳細點的
參考壹下如下參數,結合妳需要的具體要求:
1.測量範圍:0.1ppm~100%O2,
2.基本誤差:P<100ppm ±5ppm,P>100ppm ±3%;
3.零點和量程漂移:P<100ppm±5ppm/24h P>100ppm±3%/24h;
4.重復性誤差:P<100ppm±2.5ppm,P>100ppm±1.5%;
5.時間常數:<10s;
6.影響誤差;a.電壓波動:220v±10%,P<100ppm±2.5ppm,P>100ppm±1.5%;b.頻率波動:50Hz±10%,P<100ppm±2.5ppm,P>100ppm±1.5%;
c.環境溫度:0~40°C,P<100ppm±7.5ppm,P>100ppm±4.5%;d.流量波動:p<10ppm 200ml/min±5ml/min <±5ppm; p<100ppm 200ml/min±10ml/min <2.5ppm; p>100ppm 200ml/min±50ml/min <3%;
7.信號輸出:0~10mA,4~20mA,分六檔,可預置。Ⅰ.1~10ppm,Ⅱ.0~100ppm,Ⅲ.0~1000ppm,Ⅵ.0~1%,Ⅴ.0~10%,Ⅵ.0~100%;8.溫控溫度:750°C±3°C;9.預熱溫度:1h;10.功耗:50W,最大250W;11.外型、開孔尺寸及重量:300×140±350mm(寬×高×深) 7.3Kg;開孔:312±1×142±1mm(寬×高)